Description
Fluke 568Ex Intrinsically Safe Infrared Thermometer
The Fluke 568Ex Intrinsically Safe Infrared Thermometer is the one product you can use in Class I Div 1 and Div 2 or Zone 1 and 2 hazardous environments anywhere in the world. Whether you work in petroleum, chemical, oil and gas, or pharmaceutical environments, the 568 Ex allows you to carry the most trusted name in test tools into most Ex rated areas all around the globe.
With a straight-forward user interface and soft-key menus, the Fluke 568 Ex makes even complex measurements easy. Quickly navigate and adjust emissivity, save data or turn on and off alarms, with just a few pushes of a button. All in a single intrinsically safe tool certified by major rating bodies from around the world.
Fluke 568Ex Technical Specifications
Infrared Temperature Range | -40 °C to 800 °C (-40 °F to 1472 °F) |
Infrared Accuracy | < 0 °C (32 °F): ± (1.0 °C (± 2.0 °F) 0.1°/1 °C or °F); > 0 °C (32 °F): ± 1 % or ± 1.0 °C (± 2.0 °F), whichever is greater |
Display Resolution | 0.1 °C / 0.1 °F |
Infrared Spectral Response | 8 µm to 14 µm |
Infrared Response Time | < 500 msec |
Thermocouple Type-K Input Temperature Range | -270 °C to 1372 °C (-454 °F to 2501 °F) |
Thermocouple Type-K Input Accuracy | -270 to -40°C: ±(1°C 0.2°/0.1°C) (-454°F to -40°F: ±2°F 0.2°/1°F -40 to 1372°C: ±1% or 1°C (-40°F to 2501°F: ±1% or 2°F), whichever is greater |
D:S (Distance to Measurement Spot Size | 50:1 |
Laser Sighting | Single-Point Laser <1mw Ouput Class 2 (II) Operation, 630nm to 670nm |
Minimum Spot Size | 19mm (0.75″) |
Emissivity Adjustment | By Built-in Table of Common Materials or Digitally Adjustable from 0.10 to 1.00 by 0.01 |
Hi/Low Alarms | Audible and Two-Colour Visual |
MIN/MAX/AVG/DIFF | Yes |
Display | Dot Matrix 98 x 96 Pixels with Function Menus |
Backlight | Two Levels, Normal and Extra Bright for Darkened Environments |
Trigger Lock | Yes |
Switchable Celsius and Fahrenheit | Yes |
Power | 2 AA/LR6 Batteries and USB when used with a PC |
Battery Life | If used continuously: laser and backlight on, 12 hours: laser and backlight off, 100 hours |
Operating Temperature | 0 to 50°C (32°F to 122°F) |
Storage Temperature | -20 to 60°C (-40 to 140°F) |
Bead Thermocouple Type K Range | -40 to 260°C (-40 to 500°F) |
Bead Thermocouple Type K Accuracy | ±1.1°C (2.0°F) from 0°C to 260°C (32°F to 500°F), typically within 1.1°C (2.0°F) from -40°C 0°C (-40°F to 32°F) |
Safety Certifications | |
Agency | Safety Rating |
ATEX/ IECEx | Zone 1 and 2 IECEx EPS 13.0006X Ex ia IIC T4 Gb 0°C ≤ Ta ≤ 50°C EPS 13 ATEX 1.525 X II 2G Ex ia IIC T4 Gb |
NEC-500 / NEC-505 | Class I Division 1 and 2 Class I, Division 1, Groups ABCD T4 Class I, Division 2, Groups ABCD T4 Class I, Zone 1, AEx ia IIC T4 Ex ia IIC T4 0°C ≤ Ta ≤ 50°C |
GOST | Zone 1 and 2 POCC DE.ГБ05.B Ex ia IIC T4 Gb X ОТ 0°C ДО 50°C EPS 13 ATEX 1 525 X II 2G Ex ia IIC T4 Gb 0°C ≤ Ta ≤ 50°C |
PCEC | Zone 1 and 2 PCEC Ex ia IIC T4 Gb CE13. EPS 13 ATEX 1 525 X II 2G Ex ia IIC T4 Gb 0°C ≤ Ta ≤ 50°C |
INMETRO | Zone 1 and 2 IEx 13.0122X Ex ia IIC T4 Gb EPS 13 ATEX 1 525 X II 2G Ex ia IIC T4 Gb 0°C ≤ Ta ≤ 50°C |